U23分析儀器故障解析(中英文對照)

發(fā)布時間:2022-01-26 11:33  

【廣告】

fault message(故障信息) possible causes(可能的原因) remedy(補(bǔ)救辦法)

flow too low during measuring (在測量流量太低)

1:sample gas blocked or leaky(樣品氣堵塞或泄漏)

A:clean or replace the blocked parts(hose filter etc) if the problem is still present(清潔或更換堵塞的配件(軟管過濾器等)如果問題仍然存在)

2:pump not running(泵沒有運(yùn)行)

A:switch the pump on(開關(guān)泵)

B:pump capacity too low(泵的容量太低)

C:increase the pump capacity as described in section(提高泵的容量中所描述的部分)

D:pump faulty(泵故障)

no data for temperature compensation(沒有數(shù)據(jù)進(jìn)行溫度補(bǔ)償)

1:temperature compensation not completed successfully(溫度補(bǔ)償沒有成功完成)

2:new components have been loaded(新的組件被加載)

3:EEPROM has been initialized(EE PROM已初始化)flow too low during autocal(在autocal流量太低)

A:cal gas path blocked or leaky(校準(zhǔn)氣路堵塞或泄漏)

B: clean or replace the blocked parts (hose filter etc) if the problem is still present(清潔或更換堵塞的配件(軟管過濾器等)如果問題仍然存在)

conc of O2 too low during autocal (濃度的氧氣在autocal太低)

1:O2 sensor faulty or old(氧傳感器的故障或老化)

A:replace the O2 sensor as described in section(更換氧傳感器中描述的部分)meas value display( 測量值顯示)

2:O2 sensor zero not calibrated(氧傳感器的零點(diǎn)未校準(zhǔn))

calibrate the zero of the o2 sensor as described in section(校準(zhǔn)零的氧氣傳感器中描述的部分)fault at analog output(故障模擬輸出)

A:component could not be initialized when switching on(組件無法初始化時開關(guān))

B:limits were violated when calibrating the analog section(限制受到侵犯時,校準(zhǔn)模擬部分)general fault of all ir channels mv-c1/2/3 1:chopper faulty(斬波器故障)beyond tolerance(一般故障的紅外通道m(xù)v-c1 / 2 / 3超出公差)

IR source faulty(紅外源錯誤)fault of addresses for ir channels(故障紅外通道地址)

1:plug-in jumper on detector for detection of components not ok(插件在探測器檢測元件不好)

2:the cable of the detector has no contact(電纜探測器沒有接觸)